print

Cryomagnet 139OXHV40

10T vertical asymmetric dedicated to D3 - stick #2 is for TU experiments

Type

ILL VTI

Tmin

1.5K

Tmax

300K

Instrument

D3

Sensor Regul.

Cernox CX1050

Serial number

X05167

Calibration

OIS

Interpol. Table

Download (txt - 5.67 Ki)

Sensor Samp. #1

Cernox CX1050

Serial number

X48362

Calibration

ILL #2007-9

Interpol. Table

Download (txt - 5.70 Ki)

Sensor Samp. #2

Cernox CX1050

Serial number

X48295

Calibration

ILL #2007-7

Interpol. Table

Download (txt - 5.67 Ki)

IPS-120 Test 07 superconducting magnet parameters (Test 06 magnet system configuration set to 1)

S01

Amps/Tesla conversion

121.481 A -> 10 T

S02Power supply current limit

121.481 A

S03Lower (most negative) safe current

-156.25 A

S04Upper (most positive) safe current

156.25 A

S05Current breakpoints for rate limiting

b01: 45 A
b02: 100 A
b03: 110 A
b04: 121.481 A

S06Limiting rate for leads, magnet at zero

240 A/min everywhere

S07Limiting rate for leads, magnet at field

240 A/min everywhere

S08Limiting rate for magnet, "FASt" run

r01: 5.5 A/min between 0 and 45 A
r02: 4 A/min between 45 and 100 A
r03: 2 A/min between 100 and 110 A
r04: 0.5 A/min between 110 and 121.481 A

S09Limiting rate for magnet, "SLO " (slow) run

r01: 5.5 A/min between 0 and 45 A
r02: 4 A/min between 45 and 100 A
r03: 2 A/min between 100 and 110 A
r04: 0.5 A/min between 110 and 121.481 A

S10Software voltage limit

12.49 V

S11Magnet inductance

84.2 Henry according to Oxford Instruments

S15Switch heater current

70 mA