Cryomagnet 139OXHV40
10T vertical asymmetric dedicated to D3 - stick #2 is for TU experiments
Type | ILL VTI |
---|---|
Tmin | 1.5K |
Tmax | 300K |
Instrument |
Sensor Regul. | Cernox CX1050 |
---|---|
Serial number | X05167 |
Calibration | OIS |
Interpol. Table |
Sensor Samp. #1 | Cernox CX1050 |
---|---|
Serial number | X48362 |
Calibration | ILL #2007-9 |
Interpol. Table |
Sensor Samp. #2 | Cernox CX1050 |
---|---|
Serial number | X48295 |
Calibration | ILL #2007-7 |
Interpol. Table |
IPS-120 Test 07 superconducting magnet parameters (Test 06 magnet system configuration set to 1)
S01 | Amps/Tesla conversion | 121.481 A -> 10 T |
---|---|---|
S02 | Power supply current limit | 121.481 A |
S03 | Lower (most negative) safe current | -156.25 A |
S04 | Upper (most positive) safe current | 156.25 A |
S05 | Current breakpoints for rate limiting | b01: 45 A |
S06 | Limiting rate for leads, magnet at zero | 240 A/min everywhere |
S07 | Limiting rate for leads, magnet at field | 240 A/min everywhere |
S08 | Limiting rate for magnet, "FASt" run | r01: 5.5 A/min between 0 and 45 A |
S09 | Limiting rate for magnet, "SLO " (slow) run | r01: 5.5 A/min between 0 and 45 A |
S10 | Software voltage limit | 12.49 V |
S11 | Magnet inductance | 84.2 Henry according to Oxford Instruments |
S15 | Switch heater current | 70 mA |