Cryomagnet 239OXHV21
8T horizontal — Dual-exchanger VTI (alias Blue Charly)
Type | ILL VTI |
---|---|
Tmin | 1.5K |
Tmax | 300K |
Instrument | Pool |
Sensor Regul. | Cernox CX1050 |
---|---|
Serial number | X85861 |
Calibration | ILL #2013-2 |
Interpol. Table | Download (LK336) (txt - 5.65 Ki) |
Sensor Samp. | Cernox CX1050 |
---|---|
Serial number | X98509 |
Calibration | ILL #2017-5 |
Interpol. Table | Download (LK336) (txt - 5.65 Ki) |
Sensor Samp. #2 | Cernox CX1050 |
---|---|
Serial number | X121623 |
Calibration | ILL #2017-6 |
Interpol. Table | Download (LK336) (txt - 5.57 Ki) |
IPS-120 Test 07 superconducting magnet parameters (Test 06 magnet system configuration set to 1)
S01 | Amps/Tesla conversion | 97.094 A -> 8 T |
---|---|---|
S02 | Power supply current limit | 109.230 A |
S03 | Lower (most negative) safe current | -150.0 A |
S04 | Upper (most positive) safe current | 149.99 A |
S05 | Current breakpoints for rate limiting | b01: 60 A |
S06 | Limiting rate for leads, magnet at zero | 240 A/min everywhere |
S07 | Limiting rate for leads, magnet at field | 240 A/min everywhere |
S08 | Limiting rate for magnet, "FASt" run | r01: 12.14 A/min between 0 and 60 A |
S09 | Limiting rate for magnet, "SLO " (slow) run | r01: 5 A/min between 0 and 60 A |
S10 | Software voltage limit | 10.00 V |
S11 | Magnet inductance | 26.2 Henry according to Oxford Instruments |
S15 | Switch heater current | 65 mA |