print

Injection Sticks

Injection stick for cryostats

Special sample sticks are available from a pool on request. Sufficient advance notice to sane(at)ill.fr is required for their reservation and preparation. These equipments are developed, repaired and modified by the Advanced Neutron Environment team.

These sample sticks are provided for carrying out gas sorption analysis experiments. The most recent ones feature two injection lines installed in a vacuum chamber with a Thermocoax heater. This avoids the cooling of the injected gas at the level of the nitrogen shield. A permanent flow can be maintained using the two independent capillaries. At present, the maximum injection pressure is 80 bar except for the stick 184ILHP49.

Code

Temperature
Range (K)
Maximum
Pressure (bar)
Sample Space
 Ø x height (mm)

Type, Comment

177ILGI49

2.6 - 3203Ø49

double-way, for PT cooler

184ILHP49

1.5 - 320700Ø49

single capillary with Thermocoax heater, for Orange crysotat

197ILGI49

1.5 - 320200Ø49

capillaries isolated in vacuum and attached to Thermocoax heater, for Orange cryostat

199ILGI70

1.5 - 320200Ø70

capillaries isolated in vacuum and attached to Thermocoax heater, for Orange cryostat

200ILGI49

2.6 - 3203Ø49

capilaries attached to Thermocoax heater, for PT cooler

205ILGI49

1.5 - 3203Ø49

single capillary and Quartz cell

223ILGI49

1.5 - 550200Ø49

ANSTO-HZB-ILL design

231ILGI70

1.5 - 550200Ø70

ANSTO-HZB-ILL design

242ILGI70

1.5 - 550200Ø70

ANSTO-HZB-ILL design

243ILGI49

1.5 - 550200Ø49

ANSTO-HZB-ILL design

Instruments

CodeTemperature
Range (K)
Maximum
Pressure (bar)
Sample Space
 Ø x height (mm)

Type, Comment

IN1 - Lagrange

213ILGI491.5 - 55080Ø49

ANSTO-HZB-ILL design

 213ILGI701.5 - 32080Ø70

ANSTO-HZB-ILL design

IN8

237ILGI701.5 - 32080Ø70

ANSTO-HZB-ILL design

IN11

236ILGI491.5 - 3203Ø49

single capillary