Humidity Generator
The Humidity Generator is designed for fast and precise humidity control for small chambers. Generation of the gas flow directly at the target volume ensures fast control response time even for large changes in relative humidity. The humidity is generated by circulating nitrogen gas inside thermalised H2O and D2O reservoirs. It is also possible to mix nitrogen with an optional gas circulating in an optional liquid. In this case, the user selects the nature of the gas from a long list of gases already implemented in the mass-flow controllers.
The generator is operated by a Process Logic Controller M251 from Schneider and connected to a thermostatic bath regulating the base temperature of the Peltier elements controlling the temperature of the reservoirs. From the temperature and humidity setpoints, the total flow rate and the relative flow rates of D2O and optional liquid/gas provided by the user, the PLC determines the appropriate temperatures of the 8 reservoirs and the flow rates of the 8 mass-flow controllers. The touch screen graphical user interface displays the evolution of the physical parameters, the actual operated humidity chamber and other parameters.
As startup, the controller is in local mode (i.e. not accessible remotely) and waits for its connection to a nitrogen line or bottle. It checks the levels of liquids (H2O and D20) and proceed a refill if needed. Once feed with nitrogen gas, the controller is ready for operation and allows operations which depend on the availability of liquids and optional gas.
PLC Registers (Schneider M251)
Address | Access | Type | Description |
---|---|---|---|
0 | Read | Real | Temperature setpoint (°C) |
2 | Read | Real | Relative humidity setpoint (%RH) |
4 | Read | Real | Total flow-rate setpoint (sccm) |
6 | Read | Real | D2O flow ratio setpoint (%) |
8 | Read | Real | Optional flow ratio setpoint (%) |
10 | Read/Write | Real | Requested temperature setpoint (°C) |
12 | Read/Write | Real | Requested relative Humidity setpoint (%) |
14 | Read/Write | Real | Requested total flow setpoint (sccm) |
16 | Read/Write | Real | Requested D2O flow ratio setpoint (%) |
18 | Read/Write | Real | Requested optional flow ratio setpoint (%) |
20 | Read | Real | Sample chamber temperature (°C) |
22 | Read | Real | Sample chamber therm. bath temperature (°C) |
24 | Read/Write | Real | Requested sample chamber therm. bath temperature setpoint (°C) |
26 | Read | Real | MFC #1 flow-rate of N2 dry line (sccm) |
28 | Read | Real | MFC #2 flow-rate of N2 dry line (sccm) |
30 | Read | Real | MFC #3 flow-rate of H2O saturated line (sccm) |
32 | Read | Real | MFC #4 flow-rate of H2O saturated line (sccm) |
34 | Read | Real | MFC #5 flow-rate of D2O saturated line (sccm) |
36 | Read | Real | MFC #6 flow-rate of D2O saturated line (sccm) |
38 | Read | Real | MFC #7 flow-rate of optional line (sccm) |
40 | Read | Real | MFC #8 flow-rate of optional line (sccm) |
42 | Read | Real | Temperature of H2O bath #1 (°C) |
44 | Read | Real | Temperature of H2O bath #2 (°C) |
46 | Read | Real | Temperature of D2O bath #3 (°C) |
48 | Read | Real | Temperature of D2O bath #4 (°C) |
50 | Read | Real | Temperature of optional line bath #5 (°C) |
52 | Read | Real | Temperature of optional line bath #6 (°C) |
54 | Read | Real | Heater temperature of N2 dry line (°C) |
56 | Read | Real | Heater temperature of optional line (°C) |
58 | Read | Real | Pressure in H20 saturated line (bar) |
60 | Read | Real | Pressure in D20 saturated line (bar) |
62 | Read | Real | Pressure in optional wet line (bar) |
128 | Read/Write | Unsigned Integer | NOMAD Control — Year |
129 | Read/Write | Unsigned Integer | NOMAD Control — Month |
130 | Read/Write | Unsigned Integer | NOMAD Control — Day |
131 | Read/Write | Unsigned Integer | NOMAD Control — Hour |
132 | Read/Write | Unsigned Integer | NOMAD Control — Minute |
133 | Read/Write | Unsigned Integer | NOMAD Control — Second |
200 | Read | Boolean (bit 0) | Optional gas source (0:wet, 1:dry) |
Read | Boolean (bit 1) | Nitrogen gas source available | |
Read | Boolean (bit 2) | Optional gas source available | |
Read | Boolean (bit 3) | Optional gas wet line in use (calculated RH not reliable) | |
Read | Boolean (bit 4) | Regulation alarm | |
Read | Boolean (bit 5) | Regulation default | |
Read | Boolean (bit 6) | NOMAD Control — Setpoints discordance | |
Read/Write | Boolean (bit 7) | NOMAD Control — Select optional gas line mode (0:wet, 1:dry) | |
Read | Boolean (bit 8) | — | |
Read | Boolean (bit 9) | NOMAD Control — Idle | |
Read | Boolean (bit 10) | NOMAD Control — Running | |
Read | Boolean (bit 11) | NOMAD Control — Local (0) / Remote (1) | |
Read/Write | Boolean (bit 12) | NOMAD Control — Default Acknowledgment | |
Read/Write | Boolean (bit 13) | NOMAD Control — Update PLC clock |