T-Gradient Monochromator

The cryofurnace hosting the IN13 monochromator regulates the temperature of CaF2 crystals between -200°C and +200°C to scan the lattice parameter and thereby change the energy of the incident beam. For some measurements, the highest energy resolution is not necessary and the neutron flux at the sample can be increased by enlarging the distribution of d-spacings in the monochromator crystals. This is done by creating a thermal gradient in the direction parallel to the incident beam.

The average temperature of the CaF2 crystals is controlled by two independent aluminium faces pressed against the crystal faces. Each face is thermally anchored to a liquid nitrogen loop and heaters. The nitrogen flow-rate is regulated by a mass-flow controller and heaters are powered with standard power supplied. The temperatures of the two faces are controlled by providing two setpoints: T and ∆T. A logic process controller (Schneider M251) determines the appropriate flow-rates and regulates the heaters so that one face is stabilised at the temperature T while the other at T + ∆T.

The graphical user interface of the humain machine interface (Schenider HMI Harmony GTU) proposes 4 screens selected from a tab bar:

  • Graphs: Temperature and vacuum pressure evolution of the faces
  • Synoptic: schematic assembly of the crystals pressed by the two faces displaying all parameters
  • Configuration: for setting the mass-flow rates and max heating powers applied linearily on temperature ranges
  • History: Logs of all actions and errors since the beginning of its use

At the top of the screens, indicators are permanently displayed to report potential alarms and defaults. Data are collected and stored in a flash memory. The cooling and heating of the CaF2 crystals is automatic.

Full remote control

Computer control via Ethernet is provided for monitoring all parameters and start or stop the cooling. The communications are based on the Modbus communications protocol which is an industrial standard. We provide a Modbus XOP and routines in a Procedure file for controlling Modbus-TCP devices with Igor Pro.

DescriptionReadWriteAddress
T-face regulation temperature in °C (top)X 40 001
T-face regulation temperature in °C (bottom)X 40 003
∆T-face regulation temperature in °C (top)X 40 005
∆T-face regulation temperature in °C (bottom)X 40 007
Analysers temperature in °CX 40 009
Indium seal temperature in °C (bottom)X 40 011
Vacuum pressure in hPaX 40 013
Temperature setpoint in °C (i.e. T-face setpoint)XX40 015
∆T temperature difference setpoint in °C (∆T-face setpoint = T + ∆T)XX40 017
T-face nitrogen flow-rate setpoint in L/minX 40 019
∆T-face nitrogen flow-rate setpoint in L/minX 40 021
T-face mass-flow controller temperature in °CX 40 023
∆T-face mass-flow controller temperature in °CX 40 025
NOMAD tolerance in °CXX40 027
T-face averaged temperature in °CX 40 029
∆T-face averaged temperature in °CX 40 031
Averaged temperature in °CX 40 033
Firmware versionX 40 035
Alarm — Insufficient liquid N2 supplyX 40 201.0
Alarm — Insufficient vacuumX 40 201.1
Alarm — CaF2 temperature too highX 40 201.2
Alarm — Mass-flow Controller #1X 40 201.3
Alarm — Mass-flow Controller #2X 40 201.4
Alarm — Mass-flow controller temperature out of rangeX 40 201.5
Fault — Insufficient liquid N2 supplyX 40 201.6
Fault — Insufficient vacuumX 40 201.7
Fault— CaF2 temperature too highX 40 201.8
Fault — Mass-flow Controller #1X 40 201.9
Fault — Mass-flow Controller #2X 40 201.10
Fault — Mass-flow controller temperature out of rangeX 40 201.11
Fault — Top indium seal too hotX 40 201.12
Fault — Bottom indium seal too hotX 40 201.13
Equipment status — Idle bitX 40 202.1
Equipment status — Running bitX 40 202.2
Equipment status — Local (0)/Remote bit (1)XX40 202.3
Equipment status — Default acknowledgement bitXX40 202.4

Please find below the release history and some instructions for updating the firmware.

Release History

  • 1.00 - Initial release for prototype testing
  • 2.00 - First release installed on IN13 in March 2023