High Temperatures
Automatic Furnace Controllers
We provide seven 2 kVA and 3.5 kVA automatic power crates for regulating the temperature in the ILL-type resistive furnaces. The operating principle is quite simple. Once the vacuum gauge, the two thermocouples and the water cooling circuit are connected, simply insert the stick in the furnace, lock it and select the desired temperature. The Eurotherm 2704 controller automatically stops heating if the sample degases too much and waits until the vacuum is good enough to continue heating.
We also provide a similar but more compact power supply for regulating the temperature in the resistive furnaces which can be mounted on the cradles (e.g. on D9). The installation of the sample requires manipulating a sphere and fragile insulators and it is highly advisable to ask the Local Contact to perform this work. Once the furnace is closed, the power supply heats the sample in the same way as described above. Remote operation is possible using the Modbus serial communication protocol (RS485 serial port or Ethernet).
15 old power racks are still available on the instruments or from the pool. These aging systems are not as safe and automatic as the racks mentioned above. Apart a few of them that are dedicated to specific furnaces (e.g. the mirror furnace), most of these power racks are exchangeable and regulated with WEST controllers. In this case, remote operation is only possible using a RS485 serial communication protocol.
Power racks assigned to the pool
Power Rack | Furnace type | Controller | Power | Comment |
---|---|---|---|---|
#02 | Standard old rack for resistive furnace | WEST 5010 | 500 VA or 2.0 kVA + Kepco power supply | Spare |
#06 | Standard old rack for resistive furnace | WEST 5010 | 2 kVA | Spare |
#08 | Standard old rack for resistive furnace | WEST 5010 | 500 VA or 3.5 kVA | Spare |
#13 | Specific to mirror furnace | WEST 5010 | 2 kVA | Equipped with thyristors |
#14 | Specific to Si window furnace (SANS) | WEST 5010 | 500 VA or 3.5 kVA | Spare |
01HT3500VA | Automatic new rack | Eurotherm 2704 | 3.5 kVA | Replace old rack #10 |
02HT2000VA | Automatic new rack | Eurotherm 2704 | 2.0 kVA | Replace old rack #03 |
03HT3500VA | Automatic new rack | Eurotherm 2704 | 3.5 kVA | Spare |
04HT3500VA | Automatic new rack | Eurotherm 2704 | 3.5 kVA | Replace old rack #04 |
05HT3500VA | Automatic new rack | Eurotherm 2704 | 3.5 kVA | Replace old rack #05 |
06HT3500VA | Automatic new rack | Eurotherm 2704 | 3.5 kVA | Replace old rack #09 |
07HT2000VA | Automatic new rack | Eurotherm 2704 | 2.0 kVA | Replace old rack #07 |
Power racks assigned to the instruments
Instrument | Power Rack | Furnace type | Controller | Power | Comment |
---|---|---|---|---|---|
IN16B | #01 | Standard old rack for resistive furnace | WEST 5010 | 2.0 kVA | Generally used with 01TL13AN60 |
IN5, IN6 | #03 | Standard old rack for resistive furnace | WEST 5010 | 3.5 kVA | Generally used with 03TL17AN50 and 03TL08AV70 |
D20 | #11 | Specific to Si window furnace (SANS) | WEST 5010 | 500 VA or 2.0 kVA | Generally used with 11BD13AV20 |
D4 | #12 | Standard old rack for resistive furnace | WEST 5010 | 500 VA or 2.0 kVA | Generally used with 12BD14V20 |
IN1-Lagrange | #15 | Standard old rack for resistive furnace | WEST 5010 | 3.5 kVA | Generally used with 15TL13AN60 |
D2B | #16 | Specific rack with Eurotherm thyristors | WEST ? | 1.0 kVA | Dedicated to 16TL05A12 |
D1B | #17 | Standard old rack for resistive furnace | WEST ? | 3.5 kVA | Generally used with17TL13AV20 |
D9 | #21 | Specific rack for compact furnace | Eurotherm 2704 | Kepco power supply i.e. 100V-2.5A | Generally used with 21BD11AV10 |
IN11 | 18BD08A60 | Specific rack | - | ? kVA | Dedicated to 18BD08A60 |
IN15 | 22BD05SO30 | Specific rack | - | ? kVA | Dedicated to 22BD05SO30 |