Adsorption Troughs
The 8 to 12 adsorption troughs are controlled with a Process Logic Controller M251 from Schneider. This PLC reads the analog signals returned by the surface tensiometers. The temperature of the liquid interfaces are controlled with a thermostatic bath and monitored with compact pyrometers. A gas can be circulated above the interface via a flow-rate controller. In addition, the PLC controls the heating of the neutron and laser windows to prevent condensation. All these devices are controlled from the touch panel of the PLC which is also used to control the Langmuir Trough.
ModBus registers
Address | Access | Type | Description |
---|---|---|---|
40001 | Read | Real | A1 — Surface tension signal x (V) |
40003 | Read | Real | A1 — Coefficient a |
40005 | Read | Real | A1 — Offset due to paper probe b |
40007 | Read | Real | A1 — Constant c for zeroing the sensor (relative measurement) |
40009 | Read | Real | A1 — Surface tension a x + b - c (mN/m) |
40011 | Read | Real | A3 — Surface tension signal x (V) |
40013 | Read | Real | A3 — Coefficient a |
40015 | Read | Real | A3 — Offset due to paper probe b |
40017 | Read | Real | A3 — Constant c for zeroing the sensor (relative measurement) |
40019 | Read | Real | A3 — Surface tension a x + b - c (mN/m) |
40021 | Read | Real | B1 — Surface tension signal x (V) |
40023 | Read | Real | B1 — Coefficient a |
40025 | Read | Real | B1 — Offset due to paper probe b |
40027 | Read | Real | B1 — Constant c for zeroing the sensor (relative measurement) |
40029 | Read | Real | B1 — Surface tension a x + b - c (mN/m) |
40031 | Read | Real | B3 — Surface tension signal x (V) |
40033 | Read | Real | B3 — Coefficient a |
40035 | Read | Real | B3 — Offset due to paper probe b |
40037 | Read | Real | B3 — Constant c for zeroing the sensor (relative measurement) |
40039 | Read | Real | B3 — Surface tension a x + b - c (mN/m) |
40041 | Read | Real | C1 — Surface tension signal x (V) |
40043 | Read | Real | C1 — Coefficient a |
40045 | Read | Real | C1 — Offset due to paper probe b |
40047 | Read | Real | C1 — Constant c for zeroing the sensor (relative measurement) |
40049 | Read | Real | C1 — Surface tension a x + b - c (mN/m) |
40051 | Read | Real | C3 — Surface tension signal x (V) |
40053 | Read | Real | C3 — Coefficient a |
40055 | Read | Real | C3 — Offset due to paper probe b |
40057 | Read | Real | C3 — Constant c for zeroing the sensor (relative measurement) |
40059 | Read | Real | C3 — Surface tension a x + b - c (mN/m) |
40061 | Read | Real | D1 — Surface tension signal x (V) |
40063 | Read | Real | D1 — Coefficient a |
40065 | Read | Real | D1 — Offset due to paper probe b |
40067 | Read | Real | D1 — Constant c for zeroing the sensor (relative measurement) |
40069 | Read | Real | D1 — Surface tension a x + b - c (mN/m) |
40071 | Read | Real | D3 — Surface tension signal x (V) |
40073 | Read | Real | D3 — Coefficient a |
40075 | Read | Real | D3 — Offset due to paper probe b |
40077 | Read | Real | D3 — Constant c for zeroing the sensor (relative measurement) |
40079 | Read | Real | D3 — Surface tension a x + b - c (mN/m) |
40081 | Read | Real | Temperature setpoint (°C) |
40083 | Read | Real | A+B — Temperature (°C) |
40085 | Read | Real | C+D — Temperature (°C) |
40087 | Read | Real | A+B — Gas flow rate (l/min) |
40089 | Read | Real | A+B — Gas flow rate setpoint (l/min) |
40091 | Read | Real | C+D — Gas flow rate (l/min) |
40093 | Read | Real | C+D — Gas flow rate setpoint (l/min) |
40095 | Read | Real | A1 — Trough surface (mm2), equal to 0 if unused |
40097 | Read | Real | A2 — Trough surface (mm2), equal to 0 if unused |
40099 | Read | Real | A3 — Trough surface (mm2), equal to 0 if unused |
40101 | Read | Real | B1 — Trough surface (mm2), equal to 0 if unused |
40103 | Read | Real | B2 — Trough surface (mm2), equal to 0 if unused |
40105 | Read | Real | B3 — Trough surface (mm2), equal to 0 if unused |
40107 | Read | Real | C1 — Trough surface (mm2), equal to 0 if unused |
40109 | Read | Real | C2 — Trough surface (mm2), equal to 0 if unused |
40111 | Read | Real | C3 — Trough surface (mm2), equal to 0 if unused |
40113 | Read | Real | D1 — Trough surface (mm2), equal to 0 if unused |
40115 | Read | Real | D2 — Trough surface (mm2), equal to 0 if unused |
40117 | Read | Real | D3 — Trough surface (mm2), equal to 0 if unused |
40119 | Read/Write | Real | NOMAD Control — Temperature setpoint (°C) |
40121 | Read/Write | Real | NOMAD Control — A+B Gas flow rate setpoint (l/min) |
40123 | Read/Write | Real | NOMAD Control — C+D Gas flow rate setpoint (l/min) |
40125 | Read/Write | Real | NOMAD Control — Temperature tolerance (°C) |
40127 | Read/Write | Real | NOMAD Control — Gas flow rate tolerance (l/min) |
40129 | Read/Write | Integer | NOMAD Control — Year |
40131 | Read/Write | Integer | NOMAD Control — Month |
40133 | Read/Write | Integer | NOMAD Control — Date |
40135 | Read/Write | Integer | NOMAD Control — Hour |
40137 | Read/Write | Integer | NOMAD Control — Minute |
40139 | Read/Write | Integer | NOMAD Control — Second |
40201 | Read | Boolean (bit 0) | A+B — Temperature alarm (yellow: out of tolerance) |
40201 | Read | Boolean (bit 1) | C+D — Temperature alarm (yellow: out of tolerance) |
40201 | Read | Boolean (bit 2) | Temperature default (red) |
40201 | Read | Boolean (bit 3) | A+B — Gas flow rate alarm (yellow: out of tolerance) |
40201 | Read | Boolean (bit 4) | C+D — Gas flow rate alarm (yellow: out of tolerance) |
40201 | Read | Boolean (bit 5) | A+B — Gas flow rate default (red: out of gas) |
40201 | Read | Boolean (bit 6) | C+D — Gas flow rate default (red: out of gas) |
40201 | Read | Boolean (bit 7) | Condensation default (red: lid unplugged) |
40202 | Read | Boolean (bit 0) | NOMAD Control — Thermostatic bath default (red: no communication) |
40202 | Read | Boolean (bit 1) | NOMAD Control — Idle |
40202 | Read | Boolean (bit 2) | NOMAD Control — Running |
40202 | Read | Boolean (bit 3) | NOMAD Control — Local (0) / Remote (1) |
40202 | Read/Write | Boolean (bit 4) | NOMAD Control — Default acknowledgment |
40202 | Read/Write | Boolean (bit 5) | NOMAD Control — For resetting the PLC clock |