Cryomagnet 203OXHV39
10T vertical asymmetric with recondensation unit
The magnetic field homogeneity is 0.5% in a Ø10mm sphere centered on the beam axis. The maximum sample volume is Ø38 x 30 mm with ±5° vertical access.
| Type | ILL VTI |
|---|---|
Tmin |
1.5K |
Tmax |
300K |
Instrument |
| Sensor Regul. | Cernox CX1050 |
|---|---|
Serial number |
X48345 |
Calibration |
ILL 2007-9 |
Interpol. Table |
| Sensor Samp. | Cernox CX1050 |
|---|---|
Serial number |
X48346 |
Calibration |
ILL 2007-9 |
Interpol. Table |
| Sensor Magnet | - |
|---|---|
Serial number |
- |
Calibration |
- |
Interpol. Table |
- |
IPS-120 Test 07 superconducting magnet parameters (Test 06 magnet system configuration set to 1)
| S01 | Amps/Tesla conversion | 112.59 A -> 10 T |
|---|---|---|
S02 |
Power supply current limit |
112.59 A |
S03 |
Lower (most negative) safe current |
-112.59 A |
S04 |
Upper (most positive) safe current |
112.59 A |
S05 |
Current breakpoints for rate limiting |
b01: 80 A |
S06 |
Limiting rate for leads, magnet at zero |
240 A/min everywhere |
S07 |
Limiting rate for leads, magnet at field |
120 A/min everywhere |
S08 |
Limiting rate for magnet, "FASt" run |
r01: 3 A/min between 0 and 80 A |
S09 |
Limiting rate for magnet, "SLO " (slow) run |
r01: 3 A/min between 0 and 80 A |
S10 |
Software voltage limit |
12.49 V |
S11 |
Magnet inductance |
119.3 Henry according to Oxford Instruments |
S15 |
Switch heater current |
60 mA |