Cryomagnet 158OXHV26
15T vertical symmetric
The magnetic field homogeneity is 0.5% in a Ø10mm sphere centered on the beam axis. The beam opening is Ø20mm. The standard sample stick can be inserted so that its extremity reaches the beam height.
| Type | ILL VTI |
|---|---|
Tmin |
1.5K |
Tmax |
300K |
Instrument |
| Sensor Regul. | Cernox CX1050 |
|---|---|
Serial number |
X48398 |
Calibration |
ILL 2009-1 |
Interpol. Table |
| Sensor Samp. | Cernox CX1050 |
|---|---|
Serial number |
X48399 |
Calibration |
ILL 2009-1 |
Interpol. Table |
| Sensor Magnet | - |
|---|---|
Serial number |
- |
Calibration |
- |
Interpol. Table |
- |
IPS-120 Test 07 superconducting magnet parameters (Test 06 magnet system configuration set to 1)
ATTENTION ! The maximum current set in the power supply corresponds to 15 T. The maximum allowed field is 14.9 T and we cannot garantee greater fields. The maximum allowed field is 13.5 T when the temperature of the magnet is between 2.3 and 4.2 K. Users MUST check that the temperature of the magnet is below 2.3 K before requiring fields above 13.5 T.
| S01 | Amps/Tesla conversion | 115.336 A/T |
|---|---|---|
S02 |
Power supply current limit |
115.336 A |
S03 |
Lower (most negative) safe current |
-150.0 A |
S04 |
Upper (most positive) safe current |
149.99 A |
S05 |
Current breakpoints for rate limiting |
b01: 80 A |
S06 |
Limiting rate for leads, magnet at zero |
240 A/min everywhere |
S07 |
Limiting rate for leads, magnet at field |
240 A/min everywhere |
S08 |
Limiting rate for magnet, "FASt" run |
r01: 1.5 A/min between 0 and 80 A |
S09 |
Limiting rate for magnet, "SLO " (slow) run |
r01: 1.5 A/min between 0 and 80 A |
S10 |
Software voltage limit |
12.49 V |
S11 |
Magnet inductance |
194.1 Henry according to Oxford Instruments |
S15 |
Switch heater current |
68 mA |