This website is being migrated. In case you don't find the information you are looking for, the old website can be found here. Thank you for reporting any anomalies to communication@ill.eu.

Cryomagnet ORTF

2.5T horizontal - available in collaboration with T. Forgan

Type ILL VTI

Tmin

1.5K

Tmax

300K

Instrument

D33

Sensor Regul. Cernox CX1050

Serial number

X47027

Calibration

ILL 2007-3

Interpol. Table

Download

Sensor Samp. Cernox CX1050

Serial number

X47028

Calibration

ILL 2007-3

Interpol. Table

Download

IPS-120 Test 07 superconducting magnet parameters (Test 06 magnet system configuration set to 1)

S01 Amps/Tesla conversion 26.375 A -> 2.5 T

S02

Power supply current limit

26.375 A

S03

Lower (most negative) safe current

-26.375 A

S04

Upper (most positive) safe current

26.375 A

S05

Current breakpoints for rate limiting

b01: 26.375 A

S06

Limiting rate for leads, magnet at zero

240 A/min everywhere

S07

Limiting rate for leads, magnet at field

240 A/min everywhere

S08

Limiting rate for magnet, "FASt" run

r01: 0.5 A/min between 0 and 26.375 A

S09

Limiting rate for magnet, "SLO " (slow) run

r01: 0.5 A/min between 0 and 26.375 A

S10

Software voltage limit

V

S11

Magnet inductance

Henry according to

S15

Switch heater current

mA