Cryomagnet 239OXHV21
8T horizontal — Dual-exchanger VTI (alias Blue Charly)
| Type | ILL VTI |
|---|---|
Tmin |
1.5K |
Tmax |
300K |
Instrument |
| Sensor Regul. | Cernox CX1050 |
|---|---|
Serial number |
X85861 |
Calibration |
ILL #2013-2 |
Interpol. Table |
| Sensor Samp. | Cernox CX1050 |
|---|---|
Serial number |
X98509 |
Calibration |
ILL #2017-5 |
Interpol. Table |
| Sensor Samp. #2 | Cernox CX1050 |
|---|---|
Serial number |
X121623 |
Calibration |
ILL #2017-6 |
Interpol. Table |
IPS-120 Test 07 superconducting magnet parameters (Test 06 magnet system configuration set to 1)
| S01 | Amps/Tesla conversion | 97.094 A -> 8 T |
|---|---|---|
S02 |
Power supply current limit |
109.230 A |
S03 |
Lower (most negative) safe current |
-150.0 A |
S04 |
Upper (most positive) safe current |
149.99 A |
S05 |
Current breakpoints for rate limiting |
b01: 60 A |
S06 |
Limiting rate for leads, magnet at zero |
240 A/min everywhere |
S07 |
Limiting rate for leads, magnet at field |
240 A/min everywhere |
S08 |
Limiting rate for magnet, "FASt" run |
r01: 12.14 A/min between 0 and 60 A |
S09 |
Limiting rate for magnet, "SLO " (slow) run |
r01: 5 A/min between 0 and 60 A |
S10 |
Software voltage limit |
10.00 V |
S11 |
Magnet inductance |
26.2 Henry according to Oxford Instruments |
S15 |
Switch heater current |
65 mA |