Cryomagnet 134OXHV38
3.8T horizontal
| Type | ILL VTI |
|---|---|
Tmin |
1.5K |
Tmax |
300K |
Instrument |
| Sensor Regul. | Cernox CX1050 |
|---|---|
Serial number |
X177538 |
Calibration |
- |
Interpol. Table |
| Sensor Samp. | Cernox CX1050 |
|---|---|
Serial number |
X48161 |
Calibration |
ILL 2007-7 |
Interpol. Table |
IPS-120 Test 07 superconducting magnet parameters (Test 06 magnet system configuration set to 1)IPS-120 Test 07 superconducting magnet parameters (Test 06 magnet system configuration set to 1)
| S01 | Amps/Tesla conversion | 87.543 A -> 3.8 T |
|---|---|---|
S02 |
Power supply current limit |
87.543 A |
S03 |
Lower (most negative) safe current |
-150.0 A |
S04 |
Upper (most positive) safe current |
149.99 A |
S05 |
Current breakpoints for rate limiting |
b01: 74 A |
S06 |
Limiting rate for leads, magnet at zero |
240 A/min everywhere |
S07 |
Limiting rate for leads, magnet at field |
240 A/min everywhere |
S08 |
Limiting rate for magnet, "FASt" run |
r01: 6 A/min between 0 and 74 A |
S09 |
Limiting rate for magnet, "SLO " (slow) run |
r01: 6 A/min between 0 and 74 A |
S10 |
Software voltage limit |
12.49 V |
S11 |
Magnet inductance |
74.8 Henry according to Oxford Instruments |
S15 |
Switch heater current |
100 mA |