Cryomagnet 239OXHV21

8T horizontal — Dual-exchanger VTI (alias Blue Charly)

Type ILL VTI

Tmin

1.5K

Tmax

300K

Instrument

Pool

Sensor Regul. Cernox CX1050

Serial number

X85861

Calibration

ILL #2013-2

Interpol. Table

Download

Sensor Samp. Cernox CX1050

Serial number

X98509

Calibration

ILL #2017-5

Interpol. Table

Download

Sensor Samp. #2 Cernox CX1050

Serial number

X121623

Calibration

ILL #2017-6

Interpol. Table

Download

IPS-120 Test 07 superconducting magnet parameters (Test 06 magnet system configuration set to 1)

S01 Amps/Tesla conversion 97.094 A -> 8 T

S02

Power supply current limit

109.230 A

S03

Lower (most negative) safe current

-150.0 A

S04

Upper (most positive) safe current

149.99 A

S05

Current breakpoints for rate limiting    

b01: 60 A
b02: 85 A
b03: 97.094 A
 

S06

Limiting rate for leads, magnet at zero

240 A/min everywhere

S07

Limiting rate for leads, magnet at field

240 A/min everywhere

S08

Limiting rate for magnet, "FASt" run

r01: 12.14 A/min between 0 and 60 A
r02: 12.14 A/min between 60 and 85 A
r03: 12.14 A/min between 85 and 97.094 A

S09

Limiting rate for magnet, "SLO " (slow) run    

r01: 5 A/min between 0 and 60 A
r02: 2 A/min between 60 and 85 A
r03: 1 A/min between 85 and 97.094 A

S10

Software voltage limit

10.00 V

S11

Magnet inductance

26.2 Henry according to Oxford Instruments

S15

Switch heater current

65 mA