Characteristics

Characteristics

H521 guide
Peak neutron energy 11 meV / 2.8 Å
Fastest tomography 1.5 s
Wavelength Definition
Wavelength resolution DCM Δλ/λ ̴ 5%
Wavelength resolution VS Δλ/λ ̴ 16%
Sample area
Neutron flux at sample position 3×108neutrons cm-2s-1mA-1(L/D of 333)
Max. field of view 170×170 mm2
Max. spatial resolution 3 μm
Collimation
L (distance pinhole – sample) 5.5–10.5 m
D (pinhole diameter) 5, 10, 15, 23, 30, 40 mm

Neutron camera set-ups

  • Medium resolution detector with continuous adjustable FoV:
FoV [mm2] Pixel size [μm]
from 30 × 30 to 170 ×170 * from 15 to 83
  • High resolution detectors **:
FoV [mm2] Pixel size [μm]
42 × 42 21
32.5 × 32.5 16
14.6 × 14.6 7.1
8.6 × 8.6 4.2
7.3 × 7.3 3.6
5.4 × 5.4 2.6
4.1 × 4.1 1.9

* Optimal FoV above 80 × 80 mm2

** Additional configurations are currently under development and available for tests.