Cryomagnet 203OXHV39

10T vertical asymmetric with recondensation unit

The magnetic field homogeneity is 0.5% in a Ø10mm sphere centered on the beam axis. The maximum sample volume is Ø38 x 30 mm with ±5° vertical access.    

Type ILL VTI

Tmin

1.5K

Tmax

300K

Instrument

Pool

Sensor Regul. Cernox CX1050

Serial number

X48345

Calibration

ILL 2007-9

Interpol. Table

download (LK336)
download (LK340)

Sensor Samp. Cernox CX1050

Serial number

X48346

Calibration

ILL 2007-9

Interpol. Table

download (LK336)
download (LK340)

Sensor Magnet -

Serial number

-

Calibration

-

Interpol. Table

-

IPS-120 Test 07 superconducting magnet parameters (Test 06 magnet system configuration set to 1)

S01 Amps/Tesla conversion 112.59 A -> 10 T

S02

Power supply current limit

112.59 A

S03

Lower (most negative) safe current

-112.59 A

S04

Upper (most positive) safe current

112.59 A

S05

Current breakpoints for rate limiting

b01: 80 A
b02: 100 A
b03: 112.59 A

S06

Limiting rate for leads, magnet at zero

240 A/min everywhere

S07

Limiting rate for leads, magnet at field

120 A/min everywhere

S08

Limiting rate for magnet, "FASt" run

r01: 3 A/min between 0 and 80 A
r02: 2 A/min between 80 and 100 A
r03: 1 A/min between 100 and 112.59 A

S09

Limiting rate for magnet, "SLO " (slow) run

r01: 3 A/min between 0 and 80 A
r02: 2 A/min between 80 and 100 A
r03: 1 A/min between 100 and 112.59 A

S10

Software voltage limit

12.49 V

S11

Magnet inductance

119.3 Henry according to Oxford Instruments

S15

Switch heater current

60 mA