SuperADAM CRG - Advanced Reflectometer for the Analysis of Materials

Characteristics

Guide hall n°2, cold guide H521
horizontal scattering geometry
0 Å-1< Q < 2.5 Å-1
monochromator
type HOPG (002) GIC(under development)
wavelength λ 5.21 Å
resolution δλ/λ 0.005 0.035
λ/2, λ/3... filter (removable) solid state wavelength filter cooled Be
Collimation
collimation slit distance (variable) 2.5 - 4 m 3 - 5.2m
monochromator-to-sample-distance (variable) 2.8 - 5 m 3.5 - 6 m
Sample
flux (non polarized) 6×105n s-1cm-2 6×106n s-1cm-2(expected value)
flux (polarized) 3.2×105
orientation horizontal scattering plane
Sample environment
magnetic field up to 1.1T (optional 7T cryomagnet)
temperature from 2K to 400K
Detectors
3He single detector
distance to sample (variable) 1.1 - 6 m
background (beam closed) 5 cts per minute
2 dim. position sensitive3He detector
distance to sample (variable) 1.3 - 6 m
active area 30 x 30 cm2
efficiency 76 %
resolution 2.8 mm (FWHM)
Polarisation and polarisation analysis
polarisers reflection supermirror 3He spin filter
efficiency 99.8% up to 99.9% (variable)
180º spin-flippers resonance coils
analyzer reflection/transmition supermirros or3He spin filter
efficiency 99.4% (99.9% with3He spin filter)

Instrument description

The reflectometer SuperADAM is an angle dispersive fixed wavelength machine which offers two operation modes:

- high-resolution and high-polarization mode with a focussing pyrolytic graphite monochromator for studies of magnetic layered systems

- high-flux low-resolution unpolarized option for soft-matter investigations (under construction).

A two dimensional position sensitive detector allows time efficient collection of off-specular data. With a maximum scattering angle of about 85° of the detector arm Q-values up to 2.1 Å-1 can be reached for complementary measurements probing small length scales. The flexibility of the instrument permits the variation of the detector to sample and sample to monochromator distances in order to accommodate the specific requirements for incident and outgoing resolutions.

Specialised sample environment can be provided including a cryo-furnace and an electromagnet or cryomagnet.