print

Cryomagnet 283HTHV39

12T vertical asymmetric

TypeHigh-Tc magnet
Tmin1.5K
Tmax

300K

InstrumentPool

Sensor Regul.

Cernox CX1050

Serial number

X177564

Calibration

ILL #2024-1

Interpol. Table

Download (txt - 5.68 Ki)

Sensor Samp. #1

Cernox CX1050

Serial number

X177565

Calibration

ILL #2024-1

Interpol. Table

Download (txt - 5.68 Ki)

The magnet is protected by a magnet monitor which continuously checks that operating voltages and temperatures do not exceed set limits. This monitor will not allow the power supply to operate until temperatures and voltages fall below the Enable limits. After that, if voltages or temperatures rise above the Disable limits, the monitor sends a disable signal to the power supply and the operating current drops to zero to protect the magnet.

Amps/Tesla conversion18,347 A/T 
Power supply current limits-221 A+221 A
Stage 1 cooler temperature limits (T1)Enable if < 43 KDisable if > 45 K
Stage 2 cooler temperature limits (T2)Enable if < 11 KDisable if > 12 K
Coil pack A temperature limits (T3)Enable if < 13 KDisable if > 15 K
Coil pack B temperature limits (T4)Enable if < 13 KDisable if > 15 K
Interconnection temperature limits (T5)Enable if < 14 KDisable if > 16.5 K
Coil differential voltage limitsEnable if < 2 mVDisable if > 200 mV
Coil-pack A differential voltage limitsEnable if < 2 mVDisable if > 2000 mV
Coil-pack B differential voltage limitsEnable if < 2 mVDisable if > 4800 mV
Current lead differential voltage limitsEnable if < 2 mVDisable if > 5 mV