Cryomagnet CRG-CEA
12T vertical asymmetric
Sensor Samp. #1 | Cernox CX1050 |
---|---|
Serial number | X42036, X48343 |
Calibration | ILL 2007-1, 2007-9 |
Interpol. Table | Download (LK336) (txt - 5.70 Ki) Download (LK336) (txt - 5.70 Ki) |
Sensor Samp. #2 | Cernox CX1050 |
---|---|
Serial number | X87162 |
Calibration | ILL 2013-2 |
Interpol. Table | Download (LK336) (txt - 5.67 Ki) |
IPS-120 Test 07 superconducting magnet parameters (Test 06 magnet system configuration set to 1)
S01 | Amps/Tesla conversion | 120 A -> 14.688 T (98.05 -> 12 T) |
---|---|---|
S02 | Power supply current limit | + 120 A |
S03 | Lower (most negative) safe current | - 98.05 A |
S04 | Upper (most positive) safe current | 98.05 A |
S05 | Current breakpoints for rate limiting | b01: 80 A |
S06 | Limiting rate for leads, magnet at zero | 20 A/min everywhere |
S07 | Limiting rate for leads, magnet at field | 20 A/min everywhere |
S08 | Limiting rate for magnet, "FASt" run | r01: 2.5 A/min between 0 and 80 A |
S09 | Limiting rate for magnet, "SLO " (slow) run | r01: 2 A/min between 0 and 80 A |
S10 | Software voltage limit | 10 V |
S11 | Magnet inductance | 186 Henry according to Oxford Instruments |
S15 | Switch heater current | 55 mA |