PorTo

Description

PorTo is a cold neutron tomography instrument optimized for high-resolution imaging across a broad range of scientific applications. Its configuration supports both neutron radiography and tomography, with a maximum field of view of up to 80 × 80 mm².
The instrument is well suited for high-resolution studies of static samples as well as operando and in-situ experiments, accommodating laboratory-scale setups.

Applications

  • Material science and Engineering
  • Energy science and battery studies
  • Additive manufacturing and metallurgy
  • Earth and planetary Sciences
  • Fluid flow and percolation
  • Porous media research
  • Food science

Instrument layout

PorTo exploits the full intensity of the polychromatic neutron beam delivered by the H113 guide. A virtual source located 10 m upstream, combined with a set of interchangeable pinhole elements, allows collimation ratios ranging from 333 to 4000. The beam is further conditioned by a series of diaphragms along the vacuum flight tube upstream of the sample.
The imaging station is equipped with a high-precision rotation table to enable tomographic acquisition. Downstream of the sample, an indirect detector assembly records transmission images using neutron attenuation contrast.