Cryomagnet 283HTHV39

12T vertical asymmetric

TypeHigh-Tc magnet
Tmin1.5K
Tmax

300K

InstrumentPool
Sensor Regul.Cernox CX1050
Serial numberX177564
CalibrationILL #2024-1
Interpol. TableDownload (txt - 5.68 Ki)
Sensor Samp. #1Cernox CX1050
Serial numberX177565
CalibrationILL #2024-1
Interpol. TableDownload (txt - 5.68 Ki)
Sensor T2Cernox CX1050
Serial numberX48378
CalibrationILL #2025-4
Interpol. TableDownload (txt - 5.67 Ki)
Sensor T4Cernox CX1050
Serial numberX211960
CalibrationILL #2025-4
Interpol. TableDownload (txt - 5.68 Ki)
Sensor T5 thermalisationCernox CX1050
Serial numberX213165
CalibrationILL #2025-4
Interpol. TableDownload (txt - 5.68 Ki)
Sensor Stainless Steel TopCernox CX1050
Serial numberX217936
CalibrationILL #2025-4
Interpol. TableDownload (txt - 5.68 Ki)

The magnet is protected by a magnet monitor which continuously checks that operating voltages and temperatures do not exceed set limits. This monitor will not allow the power supply to operate until temperatures and voltages fall below the Enable limits. After that, if voltages or temperatures rise above the Disable limits, the monitor sends a disable signal to the power supply and the operating current drops to zero to protect the magnet.

Amps/Tesla conversion18,347 A/T 
Power supply current limits-221 A+221 A
Stage 1 cooler temperature limits (T1)Enable if < 43 KDisable if > 45 K
Stage 2 cooler temperature limits (T2)Enable if < 11 KDisable if > 12 K
Coil pack A temperature limits (T3)Enable if < 13 KDisable if > 15 K
Coil pack B temperature limits (T4)Enable if < 13 KDisable if > 15 K
Interconnection temperature limits (T5)Enable if < 14 KDisable if > 16.5 K
Coil differential voltage limitsEnable if < 2 mVDisable if > 200 mV
Coil-pack A differential voltage limitsEnable if < 2 mVDisable if > 2000 mV
Coil-pack B differential voltage limitsEnable if < 2 mVDisable if > 4800 mV
Current lead differential voltage limitsEnable if < 2 mVDisable if > 5 mV