This website is being migrated. In case you don't find the information you are looking for, the old website can be found here. Thank you for reporting any anomalies to communication@ill.eu.
Switch instrument:
Virtual tour
Tyrex-2 Polarised helium-3 filling station

Technical details

Compressor

Powerful hydraulic actuator under closed-loop control

Velocity: 30 cm/sec, final position within 3 µm precision

Polarisation Preserving Piston and Chamber: Titanium alloy, 15 cm diameter and 60 cm stroke

Total displacement: 10600 cm3, compression ratio: > 15000

Lasers

Infrared fibre laser (from ALS)

Optimised for helium gas 23S1-23P transition (1083 nm, 2 GHz bandwidth)

Power: 100 W (10 x 10 W)

Magnetic field

9 coils of 2 m diameter

Field amplitude: 10 Gauss

Homogeneity region:  diameter 1.2 m., length 2.5 m

Relative field gradient: < 3. 10-4 cm-1

Overall performance

85% polarisation in static mode

80% on neutron beam

Production rate:  2.5 bar.litre/hour

Final pressure: up to 4 bar

Helium-3 laboratory, SON group (neutron optics)

TYREX production from 2002 to 2024

The 3He neutron absorption cross-section is extremely spin dependent. The polarised gas acts as a neutron spin-filter (NSF) both for creating the polarisation of the incident beam and for analysing the polarisation after the sample.
3He has a very large 1/v absorption cross-section and only absorbs neutrons in the opposite spin state to that of the 3He nucleus. The cross-section for scattering is very small and essentially isotropic. These two features make polarised 3He NSFs a very efficient and also a very “clean” technology for neutron beam polarisation. The filters are optically neutral in contrast to other polarisation methods, which invariably modify the angular and/or wavelength distribution and can also produce spurious effects, such as off-specular or small-angle scattering (supermirrors) or multiple or higher-order Bragg scattering (polarising crystals). The beam polarisation obtained with an NSF is perfectly homogeneous over the beam profile and there is no γ-background created by the absorption of the neutrons in the NSF.

The level of gas polarisation is the critical parameter for the performance of the NSF. The Tyrex-2 central filling station typically provides 3He gas at a polarisation of about 80% on the neutron instrument. After being polarised on Tyrex-2, the 3He gas will inevitably start to depolarise with a characteristic depolarisation time T1. The main relaxation contributions typically arise from the material of the NSF cells and magnetic field gradients. Relaxation times are currently of the order of 100-300 hours, which means that the NSF gas is typically replaced on the instrument once every 1-2 days. The polarisation decays in a regular and predictable way, allowing accurate data correction.

Now available in sufficiently large quantities for several instruments to be supplied every day via the Tyrex-2 central facility, 3He NSFs are becoming an attractive solution for neutron beam polarisation over large solid angles and for the whole range of neutron wavelengths.

The Tyrex-2 filling station for nuclear polarisation of 3He gas was constructed at the ILL during the period 2019-2023 to replace the previous Tyrex station, which provided polarised 3He for ILL instruments from 2002 to 2021. The Tyrex-2 machine uses metastability exchange optical pumping for polarising the 3He gas at about 1 mbar pressure in ten optical pumping cells. The gas is then compressed up to several bars via a hydraulic titanium-alloy piston compressor. The machine can provide about 2.5 bar.l/h of 78-82% polarised gas.

Reference: Andersen K.H., Chung R., Guillard V., Humblot H., Jullien D., Lelièvre-Berna E., Petoukhov A., Tasset F., Physica B 356, 103-108 (2005). DOI: 10.1016/j.physb.2004.10.057