Cryomagnet 176ASHV49
6T vertical symmetric
Type | ILL VTI |
---|---|
Tmin | 1.5K |
Tmax | 300K |
Instrument |
Sensor Regul. | Cernox CX1020 |
---|---|
Serial number | X28899 |
Calibration | LakeShore |
Interpol. Table |
Sensor Samp. | Cernox CX1050 |
---|---|
Serial number | X144620 |
Calibration | ILL 2019-1 |
Interpol. Table |
Sensor Magnet | - |
---|---|
Serial number | - |
Calibration | - |
Interpol. Table | - |
IPS-120 Test 07 superconducting magnet parameters (Test 06 magnet system configuration set to 1)
S01 | Amps/Tesla conversion | 50.70 A -> 6 T |
---|---|---|
S02 | Power supply current limit | 50.70 A |
S03 | Lower (most negative) safe current | -50.70 A |
S04 | Upper (most positive) safe current | 50.70 A |
S05 | Current breakpoints for rate limiting | b01: 25 A |
S06 | Limiting rate for leads, magnet at zero | 240 A/min everywhere |
S07 | Limiting rate for leads, magnet at field | 120 A/min everywhere |
S08 | Limiting rate for magnet, "FASt" run | r01: 1.5 A/min between 0 and 25 A |
S09 | Limiting rate for magnet, "SLO " (slow) run | r01: 1.5 A/min between 0 and 25 A |
S10 | Software voltage limit | 10 V |
S11 | Magnet inductance | 37.4 Henry according to Oxford Instruments |
S15 | Switch heater current | 50 mA |