Cryomagnet 206OXHV26
7T vertical asymmetric - for reflectometry and SANS
Type | ILL VTI |
---|---|
Tmin | 1.5K |
Tmax | 300K |
Instrument | Pool |
Sensor Samp. #1 | Cernox CX1050 |
---|---|
Serial number | X48391 |
Calibration | ILL 2007-11 |
Interpol. Table |
Sensor Magnet | - |
---|---|
Serial number | - |
Calibration | - |
Interpol. Table | - |
Sensor Regul. | Cernox CX1050 |
---|---|
Serial number | X48390 |
Calibration | ILL 2007-11 |
Interpol. Table |
Sensor Samp. #2 | Cernox CX1050 |
---|---|
Serial number | X48356 |
Calibration | ILL 2007-9 |
Interpol. Table |
IPS-120 Test 07 superconducting magnet parameters (Test 06 magnet system configuration set to 1)
S01 | Amps/Tesla conversion | 96.79 A -> 7 T |
---|---|---|
S02 | Power supply current limit | 96.79 A |
S03 | Lower (most negative) safe current | -96.79 A |
S04 | Upper (most positive) safe current | 96.79 A |
S05 | Current breakpoints for rate limiting | b01: 50 A |
S06 | Limiting rate for leads, magnet at zero | 240 A/min everywhere |
S07 | Limiting rate for leads, magnet at field | 240 A/min everywhere |
S08 | Limiting rate for magnet, "FASt" run | r01: 3.5 A/min between 0 and 50 A |
S09 | Limiting rate for magnet, "SLO " (slow) run | r01: 3.5 A/min between 0 and 50 A |
S10 | Software voltage limit | 12.49 V |
S11 | Magnet inductance | 30.97 Henry according to Oxford Instruments |
S15 | Switch heater current | 65 mA |