Cryomagnet 158OXHV26
15T vertical symmetric
The magnetic field homogeneity is 0.5% in a Ø10mm sphere centered on the beam axis. The beam opening is Ø20mm. The standard sample stick can be inserted so that its extremity reaches the beam height.
Type | ILL VTI |
---|---|
Tmin | 1.5K |
Tmax | 300K |
Instrument | Pool |
Sensor Regul. | Cernox CX1050 |
---|---|
Serial number | X48398 |
Calibration | ILL 2009-1 |
Interpol. Table |
Sensor Samp. | Cernox CX1050 |
---|---|
Serial number | X48399 |
Calibration | ILL 2009-1 |
Interpol. Table |
Sensor Magnet | - |
---|---|
Serial number | - |
Calibration | - |
Interpol. Table | - |
IPS-120 Test 07 superconducting magnet parameters (Test 06 magnet system configuration set to 1)
ATTENTION ! The maximum current set in the power supply corresponds to 15 T. The maximum allowed field is 14.9 T and we cannot garantee greater fields. The maximum allowed field is 13.5 T when the temperature of the magnet is between 2.3 and 4.2 K. Users MUST check that the temperature of the magnet is below 2.3 K before requiring fields above 13.5 T.
S01 | Amps/Tesla conversion | 115.336 A/T |
---|---|---|
S02 | Power supply current limit | 115.336 A |
S03 | Lower (most negative) safe current | -150.0 A |
S04 | Upper (most positive) safe current | 149.99 A |
S05 | Current breakpoints for rate limiting | b01: 80 A |
S06 | Limiting rate for leads, magnet at zero | 240 A/min everywhere |
S07 | Limiting rate for leads, magnet at field | 240 A/min everywhere |
S08 | Limiting rate for magnet, "FASt" run | r01: 1.5 A/min between 0 and 80 A |
S09 | Limiting rate for magnet, "SLO " (slow) run | r01: 1.5 A/min between 0 and 80 A |
S10 | Software voltage limit | 12.49 V |
S11 | Magnet inductance | 194.1 Henry according to Oxford Instruments |
S15 | Switch heater current | 68 mA |