This website is being migrated. In case you don't find the information you are looking for, the old website can be found here. Thank you for reporting any anomalies to communication@ill.eu.

Principle

Producing cold using 3He/4He dilution

In classic dilution cryostats the liquid helium component (4.2 K) and the dilution component are combined in the same vessel. The film below [1] shows a dilution insert with the two functions separated. The insert is fitted below the liquid helium cryostat and the cold is produced in a mixing box.

In either case, gravity plays an essential role. These devices must not be overturned or even inclined.

Zero-gravity dilution

Satellites function in zero gravity, and the sample in a four-circle diffractometer like D10 rotates in every direction. This means that, if we want to cool using 3He/4He dilution, we need to find a way to obtain the mix without gravity, and that is what we see in the movie below [2].

The film shows a series of neutron images taken on Neutrograph at ILL in 2005. We can see the progressive dilution of the helium 3 droplets in helium 4 inside a capillary tube. The in situ observation of cold being produced in the D10 cryostat is possible because helium-4 is almost transparent to neutrons, whilst helium-3 is very opaque (isotopic effect).

References

  1. CD-ROM "Exploring matter with neutrons", A. Filhol (2004), Ed. iMediaSoft. Now available as a standalone application "NeutronEncyclopedia" <https://www.ill.eu/neutronencyclopedia>
  2. M. Enderle, S. Pujol, A. Hillenbach (Institut Laue-Langevin), A. Benoit (CRTBT CNRS) (2005) personal communication.