Cryomagnet 203OXHV39
10T vertical asymmetric with recondensation unit
The magnetic field homogeneity is 0.5% in a Ø10mm sphere centered on the beam axis. The maximum sample volume is Ø38 x 30 mm with ±5° vertical access.
Type | ILL VTI |
---|---|
Tmin | 1.5K |
Tmax | 300K |
Instrument | Pool |
Sensor Regul. | Cernox CX1050 |
---|---|
Serial number | X48345 |
Calibration | ILL 2007-9 |
Interpol. Table | download (LK336) (txt - 5.67 Ki) download (LK340) (txt - 5.67 Ki) |
Sensor Samp. | Cernox CX1050 |
---|---|
Serial number | X48346 |
Calibration | ILL 2007-9 |
Interpol. Table | download (LK336) (txt - 5.70 Ki) download (LK340) (txt - 5.70 Ki) |
Sensor Magnet | - |
---|---|
Serial number | - |
Calibration | - |
Interpol. Table | - |
IPS-120 Test 07 superconducting magnet parameters (Test 06 magnet system configuration set to 1)
S01 | Amps/Tesla conversion | 112.59 A -> 10 T |
---|---|---|
S02 | Power supply current limit | 112.59 A |
S03 | Lower (most negative) safe current | -112.59 A |
S04 | Upper (most positive) safe current | 112.59 A |
S05 | Current breakpoints for rate limiting | b01: 80 A |
S06 | Limiting rate for leads, magnet at zero | 240 A/min everywhere |
S07 | Limiting rate for leads, magnet at field | 120 A/min everywhere |
S08 | Limiting rate for magnet, "FASt" run | r01: 3 A/min between 0 and 80 A |
S09 | Limiting rate for magnet, "SLO " (slow) run | r01: 3 A/min between 0 and 80 A |
S10 | Software voltage limit | 12.49 V |
S11 | Magnet inductance | 119.3 Henry according to Oxford Instruments |
S15 | Switch heater current | 60 mA |