print

Cryomagnet 156OXHV49

6T vertical asymmetric - also known as 6T-CNRS

Type

ILL VTI

Tmin

1.5K

Tmax

300K

Instrument

Pool

Sensor Regul.

Cernox CX1050

Serial number

X42056

Calibration

ILL 2007-1

Interpol. Table

download (txt - 5.70 Ki)

The sample stick #2 is used to apply an electric field (0 - 20 KV AC/DC).

Sensor Samp.

Cernox CX1050

Serial number

X42057

Calibration

ILL 2007-1

Interpol. Table

download (txt - 5.70 Ki)

Sensor Samp. #2

Cernox CX1050

Serial number

X144635

Calibration

ILL 2019-1

Interpol. Table

download (txt - 5.65 Ki)

IPS-120 Test 07 superconducting magnet parameters (Test 06 magnet system configuration set to 1)

S01

Amps/Tesla conversion

101.6 A -> 6 T

S02Power supply current limit

101.6 A

S03Lower (most negative) safe current

-101.6 A

S04Upper (most positive) safe current

101.6 A

S05Current breakpoints for rate limiting

b01: 60 A
b02: 85 A
b03: 101.6 A

S06Limiting rate for leads, magnet at zero

240 A/min everywhere

S07Limiting rate for leads, magnet at field

240 A/min everywhere

S08Limiting rate for magnet, "FASt" run

r01: 5 A/min between 0 and 60 A
r02: 2.5 A/min between 60 and 85 A
r03: 1 A/min between 85 and 101.6 A

S09Limiting rate for magnet, "SLO " (slow) run

r01: 5 A/min between 0 and 60 A
r02: 2.5 A/min between 60 and 85 A
r03: 1 A/min between 85 and 101.6 A

S10Software voltage limit

10 V

S11Magnet inductance

30.7 Henry according to ILL measurements

S15Switch heater current

100 mA