NOMAD no longer captures images

If the NOMADCount command no longer captures images, it may be that the handshake between NOMAD and PSLViewer has been lost.

Try the following:

  1. Stop the NOMAD server on the LINUX machine
  2. Stop PSLViewer on the Windows 10 machine
  3. Restart PSLViewer on the Iindows 10 machine
  4. Restart NOMAD on the LINUX machine.

Nach oben

CLIP indexing not working

  1. Check the Sample-Detector distance in the Configure Laueplane window.
  2. Play with the Laue zones.  Try with more or less.  Try adjusting the trajectories.