Instruction ManualWhen installing a cryostat, cryofurnace or cryomagnet, the settings of the LakeShore 340 temperature controller must be verified and changed following the procedures detailed below. When complete, do not forget to switch on the heater by pressing the Heater Range key. Calibration curves Thermometers are installed on the heat exchanger of the variable temperature inserts and at the bottom of the sample sticks. Sometimes, there are also a few other thermometers for specific purposes. Generally, we use Cernox CX1050 sensors in cryostats and RhFe sensors in cryofurnaces. The thermometers are referenced with a serial number which is determined by visiting the Equipment section above: by clicking on the cryostat code name, you get the serial number of the thermometers. You can even download the calibration curves compatible with the LakeShore 340. By convention, the Cernox or Rhodium Iron sensor of the heat exchanger (regulation) is plugged to the Channel A of the LakeShore. The thermometer of the sample is plugged to the Channel C. In most cases, there are also Pt100 sensors attached to our equipments. They can be read with the Channels B and D respectively but they are used as spare thermometers; when the Cernox or Rhodium Iron sensors cannot be read correctly.
Zone control data entry The LakeShore 340 allows the user to establish up to 10 custom temperature zones where the controller automatically uses pre-programmed PID settings and heater range. These values have been determined at the cryogeny lab. for families of cryostats and must be entered in the controller after installing a new type of cryostat. We describe below the procedure to follow for entering these settings:
Control setup These settings normally needs to be entered only once. They do not depend on the equipment unless for very particular reasons.
Control limits The limits obviously depend on the type of equipment you are using. For orange cryostats, the maximum allowable temperature is 320K and the maximum heating power is 50W. For orange furnaces, these limits are respectively 560K and 150W. With the Kepco power amplifier connected to the LakeShore with the LKC-340 cable, the maximum heating power delivered by the LakeShore becomes 6.2mW and 62mW respectively.
|