Cryostat body wet after liquid nitrogen filling.
| Poor vacuum.
Vacuum valve not fully closed.
Vacuum leak.
| Pump the vacuum.
Close the vacuum valve.
Pump continuously and inform us.
|
Cryostat will not receive liquid helium.
| No liquid nitrogen.
Helium leak.
Contact between shields.
| Fill liquid nitrogen bath.
Pump continuously and inform us.
Return the cryostat to our lab.
|
Sample cools badly and high flow necessary to stabilise T<50K.
| Contact between shields.
| Return the cryostat to our lab. |
Cold valve blocked.
| Cryostat badly flushed.
| Reheat to 300K and inform us.
|
Temperature rises slowly after closing cold valve.
| Liquid He pressure valve open.
Leak of liquid He bath to exterior.
| Close the Liquid He pressure valve.
Return the cryostat to our lab. for repairing the leak.
|
Temperature rises rapidly after closing cold valve.
| Sample still warm.
| Wait for thermal equilibrium of sample and exchanger.
|
Cooling difficult with temperature oscillations.
| Too much exchange gas.
| Set T=10K and pump via exchange gas valve to <1mbar.
|
Sample temperature too high, temperature fluctuations or gradient.
| Sample stick too high.
| Reposition sample stick.
|
Temperature too low, fluctuations, pressure rises when heating.
| Liquid helium in the anular enclosure (tail).
| Close cold valve, set T=10K and wait until the pressure drops to < 2mbar.
|
Temperature rises for no apparent reason.
| Lack of liquid nitrogen and/or liquid helium.
| Verify the liquid helium and liquid nitrogen levels, refill.
|
Sample stick stuck in cryostat.
| Air leak into sample well.
| Set T=300K, await thermal equilibrium and remove sample stick. DO NOT FORCE AND DO NOT STAY ABOVE THE STICK.
|
T<10K, cold valve closed, pressure too high (> 5mbar)
| Warmvalve open.
| Close the warmvalve.
|