The Computing for Science (CS) group supports ILL scientists, students and visitors in a number of activities including data analysis, instrument simulation and sample simulation.

Other Light Functionalities1. Temperature controlChecks that temperature remains within specified limits. If temperature has
A warning is generated A new counting is started. The numor is thus incremented, keeping other run parameters unchanged (temporary counting, with saving). These countings may be repeated as long as the temperature is not stabilised When temperature is recovered and stable, the remaining sequence time is rescaled, so that the total execution time, including lost time, still remains close to the original execution time. During temperature changes, the sequence suspends for [wait on Reach T] before starting to count. 2. Statistics controlAll regions defined with the Regions of interest (ROIs) tool are analysed, computing the noise over signal ratio, and performing simple fits (line+gaussian and line+lorentzian, 6 parameters) on them. The normalised Chi-squared, and the estimated worse parameter uncertainty are computed. These three values give an indication about the overall statistics and if fit models give reasonable results. When all regions applied on the cumulated data (sum of repeated countings for each run) are 'good enough' (below the run options:Rel Err. specified accuracy), and if the 'Count on Statistics' has been checked, then Light ends the current counting (to save measurement time), going to the next run definition in the sequence. 3. Detector/Sample Diagnosis control(not implemented yet, Nov 2002) 4. Alert controlAll instrument control and monitoring processes may generate warning and error reports. These are mentioned into the Light Log, and raise a blinking Alert on the Light Main frame. This alert is removed when cliking on it. 5. Configuration filesLight has three types of configuration files, all with the same format.
The instrument configuration file (e.g. Config.ini) contains preferences, displays and instrument setting parameters. It may only be read/saved by in SuperUser mode. The run/sequence file is a user configuration file with only the run/sequence information. These file format is as follow: 6. How Runs/Sequences are executedRuns execution gathers a list of automatic checks and functions. Light checks that the instrument is ready to execute sequence (Idle). No other process (instrument setting, ...) should be running. 7. Automatic SetupThis procedure performs a fast automatic instrument setup.
|