guide hall nº1, cold neutron guide H112
|
|---|
beam size at sample
| 30 x 30 mm2
| sample environment
|
|---|
furnace
| 300 - 1300 K
| cryofurnace (cryostat) | 1.8 K - 560 K (300 K) | detectors
|
|---|
multidetector array (16 tubes with vertical position sensitivity)
+ 2 3He counters or up to 8 3He single counters; 2 monitors
For diffraction: not yet defined
| instrument configurations
|
|---|
| high resolution
| low resolution
| high Q | best resolution
|
|---|
monochromator
| Si(111)
| Si(111) | Si(311) | GaAS(002) option
| analysers
| Si(111) | Si(111) | Si(311) | flux at the sample
| not yet measured
| incident energy (meV)
| 2.08
| 2.08
| 7.63
| not
yet decided
| incident wavelength (Å)
| 6.271
| 6.271
| 3.275
| energy resoution (μeV)
| ∼0.35
| ∼0.9
| ∼3
| energy transfer (μeV)
| ±32
| ±32
| ±59
| Q-range (Å-1)
| 0.1 - 1.9
| 0.1 - 1.9 | 0.3 - 3.7 |
|
IN16 was on H53, IN16B will be on H112 - more efficient guide optics and band width selection
- new component: Phase Space Transformation (PST) chopper
- more solid angle
- wider dynamic range: new Doppler
- lower background : gain fom vacuum housing
|
|