SANE - Services for Advanced Neutron Environment

For more than 30 years, we develop electronics that optimize and simplify the use of our equipments. We present the most advanced instruments to monitor liquid cryogens and control cold-valves, furnaces, pressure cells...

Contact us for getting a quotation.

Back to ILL Homepage

High Magnetic Fields

Electronics measuring magnetic forces
Igor Pro template file (CTRL-Click or Right-Click to download and save this file)

Magnetic Forces Measurement Unit

For safety reasons, the magnetic forces appearing on an instrument when using a cryomagnet are measured before using a new system or after a modification of the instrument. These measurements are mandatory. The conditions in which a cryomagnet can be used are determined from these measurements and reported in the Safety section.

The vertical force is simply obtained from a dynanometer installed on the crane. The horizontal force is measured with 3 piezo sensors oriented at 120° and attached to both the cryomagnet and the sample table. The magnitude and direction of this horizontal force are calculated and displayed on a Eurotherm controller.

Please click on the left menu items to get more information on the utilization of the electronics, its specifications and a history of the revisions of the firmware.

SolidWorks representation of the magnetic forces measurement system